Catalog Number 165812
Device Problems Inflation Problem (1310); Failure to Infuse (2340)
Patient Problems No Consequences Or Impact To Patient (2199); No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete, a supplemental report will be filed.The device was not returned.
 
Event Description
It was reported the balloon did not inflate.
 
Event Description
It was reported that the balloon did not inflate.
 
Manufacturer Narrative
The device was not returned for evaluation.A potential failure mode could be ¿blocked lumen or pinched lumen¿ with a potential root cause of "wrinkled rubberize or poor inflation lumen coverage".The device history record was reviewed and found nothing that could have caused or contributed to the reported event.The instructions for use were found adequate and state the following: ¿warning: do not use ointments or lubricants having a petrolatum base.They will damage silicone.English warning: after use, this product may be a potential biohazard.Handle and dispose of in accordance with accepted medical practice and applicable laws and regulations.Units this is a single use device.Do not resterilize any portion of this device.Reuse and/or repackaging may create a risk of patient or user infection, compromise the structural integrity and/or essential material and design characteristics of the device, which may lead to device failure, and/or lead to injury, illness or death of the patient.Visually inspect the product for any imperfections or surface deterioration prior to use.¿ use luer tip syringe to inflate with stated ml of sterile water.Or ¿ for pre-filled products, remove clip and squeeze reservoir to inflate with stated ml of sterile water.For urological use only.Do not use if package is damaged.Bard, bardex, bardia, biocath and lubricath are trademarks and/or registered trademarks of c.R.Bard, inc.Single use only.Do not resterilize consult instructions for use.C.R.Bard, inc.Covington, ga 30014 usa 1 800 526 4455 www.Bardmedical.Com manufacturer: to deflate catheter balloon: gently insert a luer slip tip syringe in the catheter valve.Never use more force than is required to make the syringe ¿stick¿ in the valve.Allow the pressure within the balloon to force the plunger back and fill the syringe with water.If you notice slow or no deflation, re-seat the syringe gently.Use only gentle aspiration to encourage deflation if needed.Vigorous aspiration may collapse the inflation lumen, preventing balloon deflation.If necessary, contact adequately trained professional for assistance, as directed by hospital protocol.Should balloon rupture occur, care should be taken to assure that all balloon fragments have been removed from the patient.Female use only syringe of sterile water for balloon inflation refer to direct unit label for product content and gender specific use where applicable" h11:section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : the device was not returned.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BARDEX® ALL-SILICONE FOLEY CATHETER
Type of Device
FOLEY CATHETER
Manufacturer (Section D)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington GA 30014
MDR Report Key9512734
MDR Text Key179237937
Report Number1018233-2019-08209
Device Sequence Number1
Product Code EZL
UDI-Device Identifier00801741169786
UDI-Public(01)00801741169786
Combination Product (y/n)N
PMA/PMN Number
K040504
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,other
Type of Report Initial,Followup
Report Date 01/20/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date09/28/2023
Device Catalogue Number165812
Device Lot NumberMYDQ0143
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 12/02/2019
Initial Date FDA Received12/23/2019
Supplement Dates Manufacturer Received01/20/2020
Supplement Dates FDA Received01/20/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-