• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TRUMPF MEDIZIN SYSTEME GMBH + CO. KG TRULIGHT TRIO TUSM; SURGICAL LIGHT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TRUMPF MEDIZIN SYSTEME GMBH + CO. KG TRULIGHT TRIO TUSM; SURGICAL LIGHT Back to Search Results
Model Number 1574851
Device Problems Mechanical Problem (1384); Detachment of Device or Device Component (2907)
Patient Problem No Patient Involvement (2645)
Event Date 11/27/2019
Event Type  malfunction  
Manufacturer Narrative
A trumpf medical service technician inspected the surgical light system and found the snap ring undone.The snap ring was replaced and the light is functioning as designed.A field corrective action has been initiated by trumpf medical and reported to the fda as a result of investigations into similar events (recall number z-563-2016).The investigations have found that improper installation or servicing of the circlip in this joint can result in the slipping down or falling of the spring arm and light head components.Based on this information, no further action is required.
 
Event Description
The customer reported that the snap ring came undone.No injury reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRULIGHT TRIO TUSM
Type of Device
SURGICAL LIGHT
Manufacturer (Section D)
TRUMPF MEDIZIN SYSTEME GMBH + CO. KG
carl-zeiss strasse 7-9
saalfeld, 07318
GM  07318
Manufacturer Contact
steffen ulbrich
carl-zeiss strasse 7-9
saalfeld, 07318
GM   07318
MDR Report Key9516217
MDR Text Key202205912
Report Number9681407-2019-00059
Device Sequence Number1
Product Code FSY
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K102758
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer
Reporter Occupation Biomedical Engineer
Remedial Action Replace
Type of Report Initial
Report Date 11/27/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/24/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number1574851
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received11/27/2019
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/19/2014
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-