CLEARSTREAM TECHNOLOGIES LTD. LIFESTREAM BALLOON EXPANDABLE VASCULAR COVERED STENT Back to Search Results
Model Number LSMU1350637
Device Problems Failure to Advance (2524); Device Dislodged or Dislocated (2923)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 12/02/2019
Event Type  malfunction  
Manufacturer Narrative
As the lot number for the device was provided, a manufacturing review will be performed.The sample was not returned to the manufacturer for inspection/evaluation.Therefore, the investigation of the reported event is inconclusive.Based upon the available information, the definitive root cause for this event is unknown.The instructions for use (ifu) is adequate for the reported device/patient code(s) and provides general instructions for use, as well as warnings, precautions and potential complications associated with the device.Upon receipt of new or additional information, a follow-up report will be submitted as applicable.(expiry date 01/2022).
 
Event Description
It was reported that during a stent placement procedure in the mesenteric artery, the healthcare provider (hcp), aware of the off-label indication for use, predilated the lesion and advanced the balloon expandable vascular covered stent when it became stuck on the edge of the mesenteric artery.It was further reported that upon attempt to pull it back, the stent allegedly dislodged from the balloon and embolized down the leg to the profunda.The stent was unable to be retrieved and no further attempts will be made as it was determined that the location of the embolized stent does not pose as a health hazard to the patient.The procedure was completed by performing angioplasty.There was no reported patient injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIFESTREAM BALLOON EXPANDABLE VASCULAR COVERED STENT
Type of Device
BALLOON EXPANDABLE VASCULAR COVERED STENT
Manufacturer (Section D)
CLEARSTREAM TECHNOLOGIES LTD.
moyne upper
enniscorthy, co. wexford N A
EI  N A
Manufacturer (Section G)
CLEARSTREAM TECHNOLOGIES LTD.
moyne upper
enniscorthy, co. wexford N A
EI   N A
Manufacturer Contact
judith ludwig
1415 w. 3rd street
tempe, AZ 85281
4803032689
MDR Report Key9516524
MDR Text Key202707212
Report Number9616666-2019-00153
Device Sequence Number1
Product Code PRL
UDI-Device Identifier05391522081062
UDI-Public(01)05391522081062
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P160024
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 12/24/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberLSMU1350637
Device Catalogue NumberLSMU1350637
Device Lot NumberCMDP0211
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 12/04/2019
Initial Date FDA Received12/24/2019
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured02/13/2019
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-