• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY MITEK LLC US VAPR3 FOOTSWITCH *EA; FOOT-SWITCH, ELECTRICAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY MITEK LLC US VAPR3 FOOTSWITCH *EA; FOOT-SWITCH, ELECTRICAL Back to Search Results
Model Number 225023
Device Problem Loss of or Failure to Bond (1068)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 12/09/2019
Event Type  malfunction  
Manufacturer Narrative
Product complaint # (b)(4).Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Udi: (b)(4).
 
Event Description
It was reported by the sales rep that during a knee scope the vapr 3 footpedal was not coagulating.Another device was used to complete the procedure.No patient consequence or surgical delay was reported.
 
Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the information is unknown, not available or dof noe apfor,ation is unknown, not available or does not apply, the section/field of the form is left blank.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.H10 additional narrative: investigation summary
=
> the complaint device is not being returned, it was discarded by the customer, therefore unavailable for a physical evaluation.This complaint cannot be confirmed.With the information provided, and without the complaint device to evaluate, we cannot determine a root cause for the reported failure.The initially reported lot number was not correct as per the manufacturer.Since the device was discarded, the correct lot number is not available which precludes conducting a dhr review or a lot specific search in the complaints handling system.At this point in time, no corrective action is required, and no further action is warranted.However, depuy synthes mitek will continue to track any related complaints within this device family as a means of monitoring the extent with which this complaint is observed in the field.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Type of Device
FOOT-SWITCH, ELECTRICAL
Manufacturer (Section D)
DEPUY MITEK LLC US
325 paramount drive
raynham MA 02767
MDR Report Key9516991
MDR Text Key207510649
Report Number1221934-2019-60119
Device Sequence Number1
Product Code GEI
UDI-Device Identifier10886705009114
UDI-Public10886705009114
Combination Product (y/n)N
PMA/PMN Number
K041135
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 12/09/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/24/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number225023
Device Catalogue Number225023
Was Device Available for Evaluation? No
Date Manufacturer Received01/27/2020
Patient Sequence Number1
-
-