H10: h3,h6: the reported device, intended for use in treatment, was received for evaluation.There was no relationship found between the returned device and the reported incident.A visual inspection was performed on the product and no issue was observed.Complaint of overheating could not be reproduced.Product passed functional and high speed testing (100-10,000 rpms) for 5 minutes each in forward, reverse and oscillate directions.Unit passed functional tests on the dyonics power, dii and dii eip test control units with and without footswitch.Overheating did not occur during functional testing.The complaint was not confirmed and the root cause could not be determined since the reported malfunction could not be duplicated during the product evaluation process.A review of the device history records showed there were no indications to suggest that the product did not meet manufacturing specification or would not be able to perform as intended.A complaint history review found related failures.Please refer to the instructions for use for recommendations on proper use of the device and potential troubleshooting methods to prevent future reoccurrence of the reported event.No containment or corrective actions are recommended at this time.
|