Catalog Number 2C4711K |
Device Problems
Filling Problem (1233); Fluid/Blood Leak (1250)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 11/30/2019 |
Event Type
malfunction
|
Manufacturer Narrative
|
Should additional relevant information become available, a supplemental report will be submitted.
|
|
Event Description
|
It was reported that a small volume folfusor underinfused and leaked.This was described as ¿when the patient came back after three days, there was barely solution out of the pump." the pump was checked and the patient returned home with the same pump.The day after the patient came back and it was still not empty and it seemed that the pump had also leaked.This event was identified during patient infusion.There was no report of patient injury or medical intervention associated with this event.No additional information is available.
|
|
Manufacturer Narrative
|
The actual device was not available; however, a photograph of the sample was provided for evaluation.A visual inspection of the photograph observed white drug residue between the luer connector and the luer which indicates a leak.The exactly location of the leak could not be determined.The reported leak was verified.The cause of the leak could not be determined.The underinfusion could not be verified through photo analysis.No further testing could be performed due to nature of sample.Should additional relevant information become available, a supplemental report will be submitted.
|
|