• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BAXTER HEALTHCARE CORPORATION INFUSOR; PUMP, INFUSION, ELASTOMERIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BAXTER HEALTHCARE CORPORATION INFUSOR; PUMP, INFUSION, ELASTOMERIC Back to Search Results
Catalog Number 2C4711K
Device Problems Filling Problem (1233); Fluid/Blood Leak (1250)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 11/30/2019
Event Type  malfunction  
Manufacturer Narrative
Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
It was reported that a small volume folfusor underinfused and leaked.This was described as ¿when the patient came back after three days, there was barely solution out of the pump." the pump was checked and the patient returned home with the same pump.The day after the patient came back and it was still not empty and it seemed that the pump had also leaked.This event was identified during patient infusion.There was no report of patient injury or medical intervention associated with this event.No additional information is available.
 
Manufacturer Narrative
The actual device was not available; however, a photograph of the sample was provided for evaluation.A visual inspection of the photograph observed white drug residue between the luer connector and the luer which indicates a leak.The exactly location of the leak could not be determined.The reported leak was verified.The cause of the leak could not be determined.The underinfusion could not be verified through photo analysis.No further testing could be performed due to nature of sample.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INFUSOR
Type of Device
PUMP, INFUSION, ELASTOMERIC
Manufacturer (Section D)
BAXTER HEALTHCARE CORPORATION
deerfield IL
MDR Report Key9521880
MDR Text Key172834744
Report Number1416980-2019-07163
Device Sequence Number1
Product Code MEB
Combination Product (y/n)N
PMA/PMN Number
NA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 01/27/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/26/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Catalogue Number2C4711K
Device Lot NumberASKU
Was Device Available for Evaluation? No
Date Manufacturer Received01/09/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-