• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITHS MEDICAL ASD; INC. CLEO® 90 INFUSION SET; SET, ADMINISTRATION, INTRAVASCULAR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITHS MEDICAL ASD; INC. CLEO® 90 INFUSION SET; SET, ADMINISTRATION, INTRAVASCULAR Back to Search Results
Model Number 21-7220-24
Device Problem Disconnection (1171)
Patient Problem Cellulitis (1768)
Event Date 09/19/2019
Event Type  malfunction  
Event Description
Information was received indicating that the patient's arm was reported to catch on a smiths medical cleo® 90 infusion set, pulling out the connector and cannula.It was reported that the patient was unable to operate the connector and had difficulty disconnecting the site connector from the cannula pad.The patient experienced cellulitis at the infusion site; right lower quadrant of abdomen.Subsequently, a reported therapy was required to be initiated to resolve the cellulitis.No further adverse effects were reported.
 
Manufacturer Narrative
Other text: this mdr was generated under protocol b10009704, as a result of warning letter cms# (b)(4).A device history record (dhr) review was conducted which indicated all inspections were completed and no issues were noted during manufacture.A review of the manufacturing process for was conducted by quality engineer in order to verify that there are no situations or practices that could create the event.No discrepancies were found.No product sample was received; therefore, visual and functional testing could not be performed.The reported issue could not be confirmed as no product sample was received for evaluation.If the product is returned, the manufacturer will reopen this complaint for further investigation.No corrective actions are required since the complaint could not be confirmed since no samples or pictures were received for evaluation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CLEO® 90 INFUSION SET
Type of Device
SET, ADMINISTRATION, INTRAVASCULAR
Manufacturer (Section D)
SMITHS MEDICAL ASD; INC.
6000 nathan lane n
minneapolis, MN 55442
Manufacturer (Section G)
NULL
6000 nathan lane n
minneapolis, MN 55442
Manufacturer Contact
dave halverson
6000 nathan lane n
minneapolis,, MN 55442
7633833310
MDR Report Key9521898
MDR Text Key173723104
Report Number3012307300-2019-07322
Device Sequence Number1
Product Code FPA
UDI-Device Identifier30610586028339
UDI-Public30610586028339
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K042172
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Distributor
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 10/25/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date07/19/2023
Device Model Number21-7220-24
Device Catalogue Number21-7220-24
Device Lot Number3677257
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 12/02/2019
Initial Date FDA Received12/26/2019
Supplement Dates Manufacturer Received10/06/2022
Supplement Dates FDA Received10/25/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured09/28/2018
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-