MAUDE Adverse Event Report: MEDTRONIC NAVIGATION, INC FUSION COMPACT; EAR, NOSE AND THROAT STEREOTAXIC INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC NAVIGATION, INC FUSION COMPACT; EAR, NOSE AND THROAT STEREOTAXIC INSTRUMENT Back to Search Results
Model Number 9735602
Device Problem Imprecision (1307)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 12/09/2019
Event Type  malfunction  
Manufacturer Narrative
The unique identifier was not available at the time of reporting.No parts have been received by the manufacturer for evaluation.The manufacture date was not available at the time of reporting.If information is provided in the future, a supplemental report will be issued.
 
Event Description
Medtronic received information regarding a navigation system being used in a functional endoscopic sinus surgery (fess).It was reported that there was an alleged inaccuracy.The health care professional felt inaccurate in the frontal sinus.No inaccuracy metric available.They noticed that the non-invasive patient tracker was in the incorrect place on the 3d model so they moved it on the patient.It was unknown if re-registration was performed after moving the tracker.There was less than an hour delay in the case.No impact on patient outcome.
 
Manufacturer Narrative
H3) the software analysis was unable to determine probable cause without further information since the behavior could not be replicated.Suspecting that the patient frame moved and was not re-registered afterwards.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event Description
Medtronic received information regarding a navigation system being used in a functional endoscopic sinus surgery (fess).It was reported that there was an alleged inaccuracy.The health care professional felt inaccurate in the frontal sinus.No inaccuracy metric available.They noticed that the non-invasive patient tracker was in the incorrect place on the 3d model so they moved it on the patient.The patient was not re-registered, and the physician completed the procedure without navigation.There was less than an hour delay in the case.No impact on patient outcome.2019-dec-19 e1 rep: no additional information received 2019-dec-23 e2 rep: no additional information received.2019-dec-27 e3 rep: no additional information received.2020-feb-04 e4 rep: no additional information received.2020-feb-24 e5 rep: no additional information received.2020-3-16 703508748_e6(rep): additional information received from a manufacturer representative reported that the patient was not re -registered, and the physician completed the procedure without navigation.
 
Manufacturer Narrative
B5: additional information received from a manufacturer representative reported that the patient was not re-registered, and the physician completed the procedure without navigation.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FUSION COMPACT
Type of Device
EAR, NOSE AND THROAT STEREOTAXIC INSTRUMENT
Manufacturer (Section D)
MEDTRONIC NAVIGATION, INC
826 coal creek circle
louisville CO 80027
MDR Report Key9525696
MDR Text Key187406646
Report Number1723170-2019-06186
Device Sequence Number1
Product Code PGW
Combination Product (y/n)N
PMA/PMN Number
K153247
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup,Followup
Report Date 03/17/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number9735602
Device Catalogue Number9735602
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 12/09/2019
Initial Date FDA Received12/27/2019
Supplement Dates Manufacturer Received02/13/2020
03/16/2020
Supplement Dates FDA Received03/09/2020
03/17/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Age54 YR
Patient Weight82
-
-