Catalog Number 2C2008K
Device Problem Fluid/Blood Leak (1250)
Patient Problem No Patient Involvement (2645)
Event Date 12/02/2019
Event Type  malfunction  
Manufacturer Narrative
(b)(6).The lot 19e007 was manufactured between may 2-6, 2019.The device was received for evaluation.A visual inspection was performed, and it was noted that white striated marks resembling "crystals" were observed on the exterior surface of the bladder.The white striated marks were determined to be blooming antioxidant/ethanox which is a raw material/component of the bladder and not particulate or foreign matter.The purpose of antioxidant/ethanox is to prevent degradation of the bladder material.Blooming antioxidant/ethanox is a cosmetic condition and does not affect the functionality nor safety of the product.An ftir spectroscopy test was performed on the white striated marks and confirmed them to be ethanox/antioxidant.Functional testing was also performed, with no issues noted.The reported condition was not verified.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
It was reported that crystalized medication was observed on the surface of the balloon of a large volume infusor.Upon opening the plastic bag which contained the device, it was observed that some medication had leaked and crystalized within the device.This issue was discovered prior to patient use.There was no patient involvement.No additional information is available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INFUSOR
Type of Device
PUMP, INFUSION, ELASTOMERIC
Manufacturer (Section D)
BAXTER HEALTHCARE CORPORATION
deerfield IL
Manufacturer (Section G)
BAXTER HEALTHCARE - IRVINE
17511 armstrong avenue
building 3
irvine CA 92614
Manufacturer Contact
25212 w. illinois route 120
round lake, IL 60073
2242702068
MDR Report Key9526033
MDR Text Key172905394
Report Number1416980-2019-07174
Device Sequence Number1
Product Code MEB
UDI-Device Identifier00085412579382
UDI-Public(01)00085412579382
Combination Product (y/n)N
Reporter Country CodeEI
PMA/PMN Number
NA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Pharmacist
Type of Report Initial
Report Date 12/27/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/27/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Expiration Date05/01/2022
Device Catalogue Number2C2008K
Device Lot Number19E007
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/18/2019
Is the Reporter a Health Professional? Yes
Date Manufacturer Received12/02/2019
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-