Model Number CD3371-40QC |
Device Problems
Communication or Transmission Problem (2896); Inappropriate or Unexpected Reset (2959)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 12/17/2019 |
Event Type
malfunction
|
Manufacturer Narrative
|
The results of the investigation are inconclusive since the device was not returned for analysis.Based on the information received, the cause of the reported incident could not be conclusively determined.
|
|
Event Description
|
It was reported that the device was in back-up mode (bvvi) and the cause was unknown.No intervention has been performed at this time.The patient was stable and will continue to be monitored.
|
|
Event Description
|
Additional information received that the device was in back-up mode (bvvi) due to event queue overflow.The device was reprogrammed to resolve the event.The patient was stable and will continue to be monitored.
|
|