ABBOTT QUADRA ASSURA MP ICD; IMPLANTABLE CARDIOVERTER DEFIBRILLATOR Back to Search Results
Model Number CD3371-40QC
Device Problems Communication or Transmission Problem (2896); Inappropriate or Unexpected Reset (2959)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 12/17/2019
Event Type  malfunction  
Manufacturer Narrative
The results of the investigation are inconclusive since the device was not returned for analysis.Based on the information received, the cause of the reported incident could not be conclusively determined.
 
Event Description
It was reported that the device was in back-up mode (bvvi) and the cause was unknown.No intervention has been performed at this time.The patient was stable and will continue to be monitored.
 
Event Description
Additional information received that the device was in back-up mode (bvvi) due to event queue overflow.The device was reprogrammed to resolve the event.The patient was stable and will continue to be monitored.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
QUADRA ASSURA MP ICD
Type of Device
IMPLANTABLE CARDIOVERTER DEFIBRILLATOR
Manufacturer (Section D)
ABBOTT
645 almanor avenue
sunnyvale CA 94085
Manufacturer (Section G)
ABBOTT
645 almanor avenue
sunnyvale CA 94085
Manufacturer Contact
elizabeth boltz
15900 valley view court
sylmar, CA 91342
MDR Report Key9526267
MDR Text Key175088967
Report Number2938836-2019-17686
Device Sequence Number1
Product Code NIK
Combination Product (y/n)N
Reporter Country CodeFR
PMA/PMN Number
P030054
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date 01/14/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/27/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date04/30/2017
Device Model NumberCD3371-40QC
Device Lot NumberS000005897
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received01/02/2020
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured05/01/2015
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-