BECTON, DICKINSON AND CO. BD ULTRA FINE¿ PEN NEEDLES; HYPODERMIC SINGLE LUMEN NEEDLE Back to Search Results
Model Number 2865
Device Problems Leak/Splash (1354); Failure to Deliver (2338)
Patient Problem Underdose (2542)
Event Date 12/11/2019
Event Type  malfunction  
Manufacturer Narrative
A device evaluation and/or device history review is anticipated, but is not complete.Upon completion, a supplemental report will be filed.
 
Event Description
It was reported that bd ultra fine¿ pen needles leaked during use.This occurred on 4 occasions.The following information was provided by the initial reporter: material no.320122 batch no.9113867.It was reported that during injection insulin ran down leg and when he pulled needle out of injection site it was bent.Verbatim: from phone call on 2019-12-13 15:49:21: consumer returned call and stated that when he was doing his injection, he noticed insulin running down his leg.When he removed the needle he saw that it was bent over and realized that the insulin did not go into his skin.Consumer did not mention any of the needles breaking off.No additional information was provided.Consumer received replacement from the pharmacy.From phone call on 2019-12-13 11:54:52: follow up#2, spoke to the pharmacist, she stated she has not heard from a consumer, she will call again.Confirmed case number.Explained the pharmacist, i have processed the request to replaced the box.(bea).From phone call on 2019-12-12 10:38:56: followed up, left message with another pharmacist for updates.(bea) returned call, spoke to the pharmacist.Pharmacist stated when consumer giving injection insulin ran down the leg, when he pulled it out needle bent.He is not new to the insulin.Consumer uses new needle for his injection.She stated consumer reported it has happened with total of 4 of 5 pen needle.She also stated needle broke, she does not know if the consumer stated needle was detached or it was just bent.Provided case #, she will reach out to the consumer to contact us for updates.She has a unused pen needle from the box consumer dropped.Advised to send us 3 unused samples.Pharmacy has replaced 5mm 31g needle to see if he would like the longer needles.Explained to re check with the doctor for size, offered to send the replacement to the pharmacy (bea).Email received.Hello, we received a voicemail, stated that the customer returned a box bd nano 4mm pen needle because they are malfunctioning.
 
Manufacturer Narrative
H.6.Investigation: customer returned (95) sealed 4mm, 32g pen needles with the shelf carton from lot # 8263490.Customer states that insulin ran down the leg and the needle was bent.Thirty out of 95 returned pen needles were tested for point geometry, lube, and cannula od.All observations fall within specifications.All 30 samples were also tested and all were able to expel properly without any observed defects.Level a investigation a lot history review was carried out and no related non conformances were raised in association with this packaged lot concluding all inspections were performed per the applicable operations and met qc specifications.Root cause cannot be determined at this time as the issue is unconfirmed.
 
Event Description
It was reported that bd ultra fine¿ pen needles leaked during use.This occurred on 4 occasions.The following information was provided by the initial reporter: material no.320122 batch no.9113867 it was reported that during injection insulin ran down leg and when he pulled needle out of injection site it was bent.Verbatim: from phone call on (b)(6) 15:49:21: consumer returned call and stated that when he was doing his injection, he noticed insulin running down his leg.When he removed the needle he saw that it was bent over and realized that the insulin did not go into his skin.Consumer did not mention any of the needles breaking off.No additional information was provided.Consumer received replacement from the pharmacy.From phone call on (b)(6) 11:54:52: follow up#2, spoke to the pharmacist, she stated she has not heard from a consumer, she will call again.Confirmed case number.Explained the pharmacist, i have processed the request to replaced the box.((b)(6)) from phone call on (b)(6) 2019 10:38:56: followed up, left message with another pharmacist for updates.((b)(6)) returned call, spoke to the pharmacist.Pharmacist stated when consumer giving injection insulin ran down the leg, when he pulled it out needle bent.He is not new to the insulin.Consumer uses new needle for his injection.She stated consumer reported it has happened with total of 4 of 5 pen needle.She also stated needle broke, she does not know if the consumer stated needle was detached or it was just bent.Provided case #, she will reach out to the consumer to contact us for updates.She has a unused pen needle from the box consumer dropped.Advised to send us 3 unused samples.Pharmacy has replaced 5mm 31g needle to see if he would like the longer needles.Explained to re check with the doctor for size, offered to send the replacement to the pharmacy ((b)(6)) email received.Hello, we received a voicemail, stated that the customer returned a box bd nano 4mm pen needle because they are malfunctioning.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD ULTRA FINE¿ PEN NEEDLES
Type of Device
HYPODERMIC SINGLE LUMEN NEEDLE
Manufacturer (Section D)
BECTON, DICKINSON AND CO.
donore road
drogheda
MDR Report Key9526486
MDR Text Key191021658
Report Number9616657-2019-00388
Device Sequence Number1
Product Code FMI
UDI-Device Identifier10885403928659
UDI-Public10885403928659
Combination Product (y/n)N
PMA/PMN Number
K162516
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type other,user facility
Type of Report Initial,Followup
Report Date 02/04/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/27/2019
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model Number2865
Device Catalogue Number320122
Device Lot Number9113867
Date Manufacturer Received12/11/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Other;
-
-