• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BAXTER HEALTHCARE CORPORATION INFUSOR; PUMP, INFUSION, ELASTOMERIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BAXTER HEALTHCARE CORPORATION INFUSOR; PUMP, INFUSION, ELASTOMERIC Back to Search Results
Catalog Number 2C1063KP
Device Problem Filling Problem (1233)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 11/29/2019
Event Type  malfunction  
Manufacturer Narrative
(b)(6).Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
It was reported that a large volume infusor under infused.The issue occurred during infusion with piperacillin / tazobactam 13500mg in sodium chloride 0.9%.It was further stated that approximately 20% was reported to not have infused.There was no report of patient injury or medical intervention associated with this event.No additional information is available.
 
Manufacturer Narrative
The lot was manufactured from march 28, 2019 - march 30, 2019.The actual device was received for evaluation.Visual inspection was performed and there were no abnormalities that could have contributed to the reported condition.A functional flow rate test was performed and found to be within the product specification range.The reported condition was not verified.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INFUSOR
Type of Device
PUMP, INFUSION, ELASTOMERIC
Manufacturer (Section D)
BAXTER HEALTHCARE CORPORATION
deerfield IL
Manufacturer (Section G)
BAXTER HEALTHCARE - IRVINE
17511 armstrong avenue
building 3
irvine CA 92614
Manufacturer Contact
25212 w. illinois route 120
round lake, IL 60073
2242702068
MDR Report Key9527076
MDR Text Key172933989
Report Number1416980-2019-07197
Device Sequence Number1
Product Code MEB
UDI-Device Identifier00085412081366
UDI-Public(01)00085412081366
Combination Product (y/n)N
Reporter Country CodeAS
PMA/PMN Number
K071222
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 01/30/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/27/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Catalogue Number2C1063KP
Device Lot Number19C049
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/30/2019
Is the Reporter a Health Professional? No
Date Manufacturer Received01/07/2020
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Treatment
PIPERACILLIN / TAZOBACTAM 13500MG; SODIUM CHLORIDE 0.9%
-
-