• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BAXTER HEALTHCARE CORPORATION FOLFUSOR; PUMP, INFUSION, ELASTOMERIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BAXTER HEALTHCARE CORPORATION FOLFUSOR; PUMP, INFUSION, ELASTOMERIC Back to Search Results
Catalog Number 2C4063K
Device Problem Filling Problem (1233)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 12/02/2019
Event Type  malfunction  
Manufacturer Narrative
The device was manufactured from december 19, 2017 - december 20, 2017.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.The device was not received for evaluation; therefore, a device analysis could not be completed.Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
It was reported that a large volume folfusor underinfused while connected to a non-baxter set equipped with an anti-return valve.It was further reported that after 24 hours of diffusion, 100 ml remained in the device.The device was filled with 270ml of physiological serum (non-baxter) and 3gm of cefazoline (non-baxter).This issue was identified during patient infusion.There was no patient injury or medical intervention associated with this event.No additional information is available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FOLFUSOR
Type of Device
PUMP, INFUSION, ELASTOMERIC
Manufacturer (Section D)
BAXTER HEALTHCARE CORPORATION
deerfield IL
Manufacturer (Section G)
BAXTER HEALTHCARE - IRVINE
17511 armstrong avenue
building 3
irvine CA 92614
Manufacturer Contact
25212 w. illinois route 120
round lake, IL 60073
2242702068
MDR Report Key9527090
MDR Text Key172934341
Report Number1416980-2019-07199
Device Sequence Number1
Product Code MEB
UDI-Device Identifier00085412080161
UDI-Public(01)00085412080161
Combination Product (y/n)N
Reporter Country CodeFR
PMA/PMN Number
NA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 12/27/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/27/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Catalogue Number2C4063K
Device Lot Number17N030
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received12/04/2019
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Treatment
270ML PHYSIOLOGICAL SERUM (NON BAXTER); 3GM CEFAOZOLINE (NON BAXTER); MIDLINE CATHETER (NON BAXTER)
-
-