• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BAXTER HEALTHCARE CORPORATION EM 2400, LOAD CELL MODULE; SYSTEM/DEVICE, PHARMACY COMPOUNDING

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BAXTER HEALTHCARE CORPORATION EM 2400, LOAD CELL MODULE; SYSTEM/DEVICE, PHARMACY COMPOUNDING Back to Search Results
Catalog Number 2400L
Device Problem Break (1069)
Patient Problem No Patient Involvement (2645)
Event Date 12/04/2019
Event Type  malfunction  
Manufacturer Narrative
(b)(6).The device has been received and the evaluation is in progress.Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
It was reported that a load cell module had frayed cables; further described as "cords exposed and some metal sticking out".This was identified during an unspecified process step.There was no patient involvement.No additional information is available.
 
Manufacturer Narrative
The device was received for evaluation.A visual inspection was performed which observed the outer insulation covering of the cable was damaged/frayed exposing the ground braid and wires.No actual internal conductors exposed.Functional testing was performed which could not establish load cell communication; there was intermittent connectivity or continuity of signal wires due to the damaged frayed cord.The reported condition was verified.The cause of the condition was due to the electrical component failure.The scale cable assembly was replaced during the service refurbishment process.A service history review revealed no indication that the parts replaced caused or contributed to the reported event; the device was serviced greater than twelve (12) months.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EM 2400, LOAD CELL MODULE
Type of Device
SYSTEM/DEVICE, PHARMACY COMPOUNDING
Manufacturer (Section D)
BAXTER HEALTHCARE CORPORATION
deerfield IL
MDR Report Key9527586
MDR Text Key173041809
Report Number1416980-2019-07201
Device Sequence Number1
Product Code NEP
Combination Product (y/n)N
PMA/PMN Number
NA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 01/28/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Catalogue Number2400L
Device Lot Number36393
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/11/2019
Initial Date Manufacturer Received 12/04/2019
Initial Date FDA Received12/27/2019
Supplement Dates Manufacturer Received01/06/2020
Supplement Dates FDA Received01/28/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-