New information notes that the patient came into the office on (b)(6) 2019, with a suspected problem with the rf chip.The device was reset and then was able to communicate with the rf tower.All initial parameters were restored, and her device¿s rf capability is functioning properly.No further follow up on this matter is necessary.
|
It was reported that the device and leads were checked before and after mri.After the mri, the device was unable to be interrogated via rf.The device was able to be interrogated with the wand, and the interrogation showed normal device function.The device was re-interrogated with a new programmer and rf tower and was unable to interrogate.The following day, the patient tried to send a manual transmission from the home monitor, and it did not work.The monitor was unplugged and will try to send another manual transmission later.Device remains implanted.Patient condition is stable.
|