• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SYNCARDIA SYSTEMS, LLC SYNCARDIA FREEDOM DRIVER; EXTERNAL PNEUMATIC DRIVER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SYNCARDIA SYSTEMS, LLC SYNCARDIA FREEDOM DRIVER; EXTERNAL PNEUMATIC DRIVER Back to Search Results
Model Number 595000-001
Device Problem Output Problem (3005)
Patient Problem No Patient Involvement (2645)
Event Date 12/05/2019
Event Type  malfunction  
Manufacturer Narrative
During investigation testing, the reported high left arterial pressure (lap) values were reproduced.The root cause was determined to be a malfunction of the piston cylinder assembly (pca).Syncardia has a corrective and preventive action (capa) to investigate this issue.Syncardia has completed its evaluation of this complaint and is closing this file.(b)(4).
 
Event Description
While performing functional testing, a syncardia technician reported that the freedom driver did not maintain the proper normotensive pressure values on the patient simulator.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYNCARDIA FREEDOM DRIVER
Type of Device
EXTERNAL PNEUMATIC DRIVER
Manufacturer (Section D)
SYNCARDIA SYSTEMS, LLC
1992 e. silverlake road
tucson AZ 85713
Manufacturer (Section G)
SYNCARDIA SYSTEMS, LLC
1992 e. silverlake road
tucson AZ 85713
Manufacturer Contact
kerri hensley
1992 e. silverlake road
tucson, AZ 85713
5205451234
MDR Report Key9528616
MDR Text Key174684997
Report Number3003761017-2019-00370
Device Sequence Number1
Product Code LOZ
UDI-Device Identifier00858000003121
UDI-Public(01)00858000003121
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P030011
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type other
Reporter Occupation Administrator/Supervisor
Type of Report Initial
Report Date 12/23/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model Number595000-001
Device Catalogue Number595000-001
Was Device Available for Evaluation? Yes
Initial Date Manufacturer Received 12/06/2019
Initial Date FDA Received12/29/2019
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/27/2017
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-