• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW ORTHOPAEDICS LTD ACETLR CUP HAP 60MM W/ IMPTR; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/METAL, RESURFACING

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW ORTHOPAEDICS LTD ACETLR CUP HAP 60MM W/ IMPTR; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/METAL, RESURFACING Back to Search Results
Catalog Number 74120160
Device Problems Appropriate Term/Code Not Available (3191); Patient Device Interaction Problem (4001)
Patient Problems Pain (1994); Tissue Damage (2104); Toxicity (2333); Injury (2348); Limited Mobility Of The Implanted Joint (2671)
Event Date 04/18/2017
Event Type  Injury  
Event Description
It was reported a left hip revision due to severe pain; decreased mobility; shedding of metal debris into the bloodstream; tissue damage; and elevated cobalt and chromium levels.
 
Manufacturer Narrative
It was reported that left hip revision surgery was performed.During the revision, the bhr cup and bhr head were removed.As of today, the implanted devices, all of which were used in treatment, and additional information have been requested for this complaint but have not become available.A review of the complaint history for the bhr cup and bhr head was performed using batch numbers in search of similar recurring reports for the products during their lifetimes.No other similar complaints were identified for the bhr head.Similar complaints have been identified for the bhr cup.This will continue to be monitored.In the absence of the actual devices, the production records were reviewed for the devices reportedly involved in this incident.All the released devices involved met manufacturing specifications at the time of production.Review of the product ifu found adequate warnings and precautions in relation to the alleged failure modes.A risk management review was performed.No additional risks were identified as result of the reported event and no further actions are required at this time.The available medical documents were reviewed.Based on the limited information provided, the reported pain, limp and elevated cocr levels, and the intraoperative findings of metallosis and discolored tissue may be consistent with findings associated with metal debris.The lateral positioning of the acetabular component cannot be excluded as a result of the reported clinical symptoms and short external rotators which were noted intraoperatively to be completely destroyed.Without supporting post-primary and pre-revision radiographic images, lab reports and/or the analysis of the explanted components, the source of the reported findings cannot be confirmed, and it cannot be concluded that the reported clinical reactions were associated with a mal-performance of the implant.The patient impact beyond the revision and expected postoperative healing/pain cannot be determined.Without return of the actual devices or further information we cannot further investigate or confirm the details supplied in this complaint, and our investigation remains inconclusive.If the products or additional information become available in the future, this case will be reopened.No preventative or corrective action has been initiated as a result of this investigation.
 
Manufacturer Narrative
New information: g4, d4.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACETLR CUP HAP 60MM W/ IMPTR
Type of Device
PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/METAL, RESURFACING
Manufacturer (Section D)
SMITH & NEPHEW ORTHOPAEDICS LTD
aurora house
spa park
leamington spa CV31 3HL
UK  CV31 3HL
MDR Report Key9528620
MDR Text Key172996261
Report Number3005975929-2019-00447
Device Sequence Number1
Product Code NXT
Combination Product (y/n)N
PMA/PMN Number
P040033
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type other
Type of Report Initial,Followup,Followup
Report Date 10/22/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Lay User/Patient
Device Expiration Date09/30/2012
Device Catalogue Number74120160
Device Lot Number078534
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 12/03/2019
Initial Date FDA Received12/29/2019
Supplement Dates Manufacturer Received06/29/2020
08/18/2020
Supplement Dates FDA Received07/03/2020
10/22/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Treatment
74121154 RESURF FEM HEAD 54MM 082022.
Patient Outcome(s) Hospitalization; Required Intervention;
-
-