Catalog Number 530.610
Device Problem Detachment of Device or Device Component (2907)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 12/01/2019
Event Type  malfunction  
Manufacturer Narrative
(b)(4).As of this date, the device has not been returned for evaluation; therefore, the reported condition cannot be confirmed and/or duplicated.If additional information should become available, a supplemental medwatch report will be sent accordingly.
 
Event Description
It was reported from canada that the battery oscillator device screw cap that locked the housing in place had broken off.It was not reported if the device was used in surgery, or if there was patient involvement.It was not reported if there were any delays in the surgical procedure, or if a spare device was available for use.It was not reported if there were any injuries, medical intervention or prolonged hospitalization.The exact date of this event was unknown but was noted to have occurred in (b)(6) of 2019.All available information has been disclosed.If additional information should become available, a supplemental medwatch will be submitted accordingly.
 
Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.The actual device was returned for evaluation.During repair, an evaluation was performed and it was determined that the device failed general condition, had unknown liquid on the motor sub assembly, electronic control unit and gear components.It was further determined that the covering flange was broken off the handpiece housing, and the angle sticks and the plugs were corroded.Therefore, the reported condition was confirmed.The assignable root cause was determined to be due to user.If additional information should become available, a supplemental medwatch will be submitted accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BATTERY OSCILLATOR
Type of Device
MOTOR, SURGICAL INSTRUMENT, AC-POWERED
Manufacturer (Section D)
DEPUY SYNTHES PRODUCTS LLC
eimattstrasse 3
oberdorf 4436
SZ  4436
MDR Report Key9528654
MDR Text Key190519248
Report Number8030965-2019-71452
Device Sequence Number1
Product Code GEY
Combination Product (y/n)N
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 12/16/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number530.610
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/09/2020
Initial Date Manufacturer Received 12/16/2019
Initial Date FDA Received12/29/2019
Supplement Dates Manufacturer Received01/19/2020
Supplement Dates FDA Received01/23/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-