• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. TRAUMA IMPLANT; NAIL, FIXATION, BONE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. TRAUMA IMPLANT; NAIL, FIXATION, BONE Back to Search Results
Device Problem Loose or Intermittent Connection (1371)
Patient Problems Unspecified Infection (1930); Injury (2348)
Event Type  Injury  
Event Description
It was reported that during a literature study it was revealed that the patient the hip screw cut out the femoral head and this patient was changed to a hip prothesis.In addition to this the patient had infection which require debridement with antibiotic beads prior to hip prosthesis.
 
Manufacturer Narrative
It was reported from a literature study that the screw cut out the femoral head and the patient was changed to a hip prothesis.In addition to this the patient had infection which require debridement with antibiotic beads prior to hip prosthesis.The affected device, used in treatment, was not returned for evaluation.Therefore a product analysis could not be performed.As device information was not made available, device history record and complaint history review cannot be completed.There is no information that would suggest the device failed to meet specifications.A relationship, if any, between the device and the reported incident could not be corroborated.A clinical analysis noted that one photo with four combined x-ray images labeled figure 7 in the article was provided for review and confirms the reported failure.No individual clinical information has been provided for inclusion in this medical investigation.Without the return of the actual product involved and no patient medical records available, our investigation of this report is inconclusive.Based on this investigation, the need for corrective action is not indicated.Infection, a potential complication associated with any surgery, can occur and possible causes could include but are not limited to contamination, patient reaction, and post-operative healing issue.No further investigation is warranted for this complaint; however smith and nephew will continue to monitor for future complaints and investigate as necessary.Should the device or additional information be received, the complaint will be reopened.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRAUMA IMPLANT
Type of Device
NAIL, FIXATION, BONE
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks road
memphis TN 38116
MDR Report Key9531824
MDR Text Key173082642
Report Number1020279-2019-04659
Device Sequence Number1
Product Code JDS
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional,l
Type of Report Initial,Followup
Report Date 04/23/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/30/2019
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Was Device Available for Evaluation? No
Date Manufacturer Received04/20/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Required Intervention;
-
-