SYNCARDIA SYSTEMS, LLC SYNCARDIA FREEDOM DRIVER; EXTERNAL PNEUMATIC DRIVER Back to Search Results
Model Number 595000-001
Device Problem Display or Visual Feedback Problem (1184)
Patient Problem No Patient Involvement (2645)
Event Date 12/05/2019
Event Type  malfunction  
Manufacturer Narrative
This alleged failure mode poses a low risk to a patient because the issue was observed when the freedom driver was not supporting a patient.The freedom driver is being evaluated by syncardia.The results of the investigation will be provided in a follow-up mdr.(b)(4).
 
Event Description
While performing a routine evaluation, a syncardia technician reported that the freedom driver display won't light up after the lcd display button is depressed.
 
Manufacturer Narrative
During investigation testing, the reported issue that the freedom driver display did not light up when the display button was depressed was confirmed and reproduced.The root cause was determined to be a malfunction of the lcd printed circuit board (pcb).This issue will continue to be monitored and trended as part of the customer experience process.Syncardia has completed its evaluation of this complaint and is closing this file.(b)(4) follow-up report 1.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYNCARDIA FREEDOM DRIVER
Type of Device
EXTERNAL PNEUMATIC DRIVER
Manufacturer (Section D)
SYNCARDIA SYSTEMS, LLC
1992 e. silverlake road
tucson AZ 85713
MDR Report Key9532646
MDR Text Key174683260
Report Number3003761017-2019-00371
Device Sequence Number1
Product Code LOZ
UDI-Device Identifier00858000003121
UDI-Public(01)00858000003121
Combination Product (y/n)N
PMA/PMN Number
P030011
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type other
Type of Report Initial,Followup
Report Date 12/27/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model Number595000-001
Device Catalogue Number595000-001
Was Device Available for Evaluation? Yes
Initial Date Manufacturer Received 12/05/2019
Initial Date FDA Received12/30/2019
Supplement Dates Manufacturer Received12/05/2019
Supplement Dates FDA Received01/09/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-