Model Number N/A
Device Problem Inability to Auto-Fill (1044)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Manufacturer Narrative
The production device history record (dhr) for this intra-aortic balloon pump (iabp) is not required to be reviewed per getinge standard operating procedure since the iabp was manufactured more than a year before the date of event.A getinge service territory manager (stm) was dispatched to evaluate the iabp unit.Upon initial inspection, the stm found electrical test failure code #53 along with autofill failures.The shuttle transducer was out of calibration.Upon inspection of the shuttle transducer, it was noted there was too much vacuum grease on o-ring and inside the transducer, clogging the port.The transducer was cleaned and recalibrated.The iabp passed all testing and was cleared for clinical use.The customer has been performing maintenance on their own without any certified technicians.The stm recommend they send a technician to training or call the stm for proper pms in the future.The full name of the event site (b)(4).
 
Event Description
It was reported the cs300 intra-aortic balloon pump (iabp) had an autofill failure.It is unknown under which circumstances this event occurred or if a patient was involved; however, there wasno adverse event reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CS300
Type of Device
SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
Manufacturer (Section D)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ 07430
Manufacturer (Section G)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ 07430
Manufacturer Contact
1300 macarthur blvd
mahwah, NJ 07430
MDR Report Key9532665
MDR Text Key190337736
Report Number2249723-2019-02120
Device Sequence Number1
Product Code DSP
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K063525
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Reporter Occupation Other
Type of Report Initial
Report Date 12/30/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/30/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model NumberN/A
Device Catalogue Number0998-00-3023-53
Device Lot NumberN/A
Was Device Available for Evaluation? Yes
Date Manufacturer Received12/06/2019
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/01/2014
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse