• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL, INC.(CRM-SYLMAR) QUARTET LEAD SMALL-S, 86 CM; PERMANENT PACEMAKER ELECTRODE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL, INC.(CRM-SYLMAR) QUARTET LEAD SMALL-S, 86 CM; PERMANENT PACEMAKER ELECTRODE Back to Search Results
Model Number 1456Q/86
Device Problem Difficult to Insert (1316)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 12/12/2019
Event Type  malfunction  
Manufacturer Narrative
The results of the investigation are inconclusive since the device was not returned for analysis.Based on the information received, the cause of the reported incident could not be conclusively determined.
 
Event Description
It was reported that during implant procedure, the left ventricular lead was attempted to be repositioned, the stylet would not pass through the lead without difficulty.The lead was not used, another lead was implanted.The patient was stable.
 
Manufacturer Narrative
The damage found was sustained during the surgical procedure.The lead was otherwise normal.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
QUARTET LEAD SMALL-S, 86 CM
Type of Device
PERMANENT PACEMAKER ELECTRODE
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
elizabeth boltz
15900 valley view court
sylmar, CA 91342
MDR Report Key9533483
MDR Text Key174927777
Report Number2017865-2019-18679
Device Sequence Number1
Product Code NIK
UDI-Device Identifier05414734510158
UDI-Public05414734510158
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P030054
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 02/13/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date12/31/2021
Device Model Number1456Q/86
Device Catalogue Number1456Q/86
Device Lot NumberA000072677
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/30/2019
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 12/12/2019
Initial Date FDA Received12/30/2019
Supplement Dates Manufacturer Received02/12/2020
Supplement Dates FDA Received02/13/2020
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/25/2019
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age70 YR
Patient Weight64
-
-