• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COLOPLAST A/S TITAN SCROTAL ZERO ANG 20CM; INFLATABLE PENILE PROSTHESIS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COLOPLAST A/S TITAN SCROTAL ZERO ANG 20CM; INFLATABLE PENILE PROSTHESIS Back to Search Results
Model Number ES89202400
Device Problem Improper Flow or Infusion (2954)
Patient Problem No Information (3190)
Event Date 12/04/2019
Event Type  Injury  
Manufacturer Narrative
The lot number was reviewed for complaint trend, nonconforming report and capa review.No trends were noted.Without the benefit of examination and testing, coloplast is precluded from commenting on the condition of the device or the cause of the occurrence.Should additional facts prompt us to alter or supplement any information or conclusions contained in the original mdr or in any prior supplemental reports, a follow-up report will be submitted.
 
Event Description
According to the available information, the pump was auto inflating.The existing pump was replaced with a new titan touch pump.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TITAN SCROTAL ZERO ANG 20CM
Type of Device
INFLATABLE PENILE PROSTHESIS
Manufacturer (Section D)
COLOPLAST A/S
holtedam 1
humlebaek, 3050
DA  3050
Manufacturer (Section G)
COLOPLAST MANUFACTURING US, LLC
1601 west river road north
minneapolis MN 55411
Manufacturer Contact
lauren prioleau
1601 west river road north
minneapolis, MN 55411
6122196218
MDR Report Key9535813
MDR Text Key173226281
Report Number2125050-2019-01201
Device Sequence Number1
Product Code FHW
UDI-Device Identifier05708932487368
UDI-Public05708932487368
Combination Product (y/n)N
PMA/PMN Number
P000006
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Type of Report Initial
Report Date 12/30/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Lay User/Patient
Device Model NumberES89202400
Device Catalogue NumberES8920
Device Lot Number6949387
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 12/04/2019
Initial Date FDA Received12/31/2019
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Patient Outcome(s) Other;
Patient Age56 YR
-
-