• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: FRESENIUS HEMOCARE NETHERLANDS B.V. FRESENIUS COM.TEC AUTOMATED CELL SEPARATOR; COM.TEC PL1 RED CELL / PLASMA EXCHANGE SET

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

FRESENIUS HEMOCARE NETHERLANDS B.V. FRESENIUS COM.TEC AUTOMATED CELL SEPARATOR; COM.TEC PL1 RED CELL / PLASMA EXCHANGE SET Back to Search Results
Model Number N/A
Device Problems Insufficient Flow or Under Infusion (2182); Appropriate Term/Code Not Available (3191)
Patient Problem No Information (3190)
Event Date 12/05/2019
Event Type  malfunction  
Event Description
Patient underload.
 
Event Description
Patient underload the plasma pump tube has assembly error.No pictures or sample received.It is understood that the inlet and outlet tubes are exchanged from the plasma pump.Potential hazard / hazardous situation: tube glued into wrong port of the pump adapter by manufacturing operator.Severity of the harm: catastrophic occurrence of the harm: improbable.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FRESENIUS COM.TEC AUTOMATED CELL SEPARATOR
Type of Device
COM.TEC PL1 RED CELL / PLASMA EXCHANGE SET
Manufacturer (Section D)
FRESENIUS HEMOCARE NETHERLANDS B.V.
runde zz 41
emmer-compascuum, NL-78 81 H
NL  NL-7881 HM
MDR Report Key9537791
MDR Text Key188834024
Report Number3002807758-2019-00002
Device Sequence Number1
Product Code LKN
Combination Product (y/n)N
PMA/PMN Number
K060734
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,user facility
Type of Report Initial,Followup
Report Date 12/06/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model NumberN/A
Device Catalogue Number9400408
Device Lot NumberHET203
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Distributor Facility Aware Date12/06/2019
Event Location Hospital
Date Report to Manufacturer12/06/2019
Initial Date Manufacturer Received 12/06/2019
Initial Date FDA Received01/02/2020
Supplement Dates Manufacturer Received12/06/2019
Supplement Dates FDA Received07/20/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-