• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ADAPTEC MEDICAL DEVICES LLC 3014271001 SENSICA

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ADAPTEC MEDICAL DEVICES LLC 3014271001 SENSICA Back to Search Results
Catalog Number SCCS1001
Device Problems Incorrect, Inadequate or Imprecise Result or Readings (1535); Volume Accuracy Problem (1675)
Patient Problems No Consequences Or Impact To Patient (2199); No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete, a supplemental report will be filed.
 
Event Description
It was reported that the sensica device was not accurately recording volume.The complainant noted the device was 50-100 cc off of the accurate reading every hour.
 
Manufacturer Narrative
The reported issue was unconfirmed.The root cause of the reported issue could not be determined due to the inability to reproduce the issue.The tech could not duplicate the reported issue during testing.Unit was functionally tested and no issues were sighted.The device history record was reviewed and found nothing that could have caused or contributed to the reported event.The instructions for use were found adequate and state the following: ¿the monitor display shows how full the bag is, updating continuously along with the current hour uo measurement." h11:section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Event Description
It was reported that the sensica device was not accurately recording volume.The complainant noted the device was 50-100cc off of the accurate reading every hour.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SENSICA
Type of Device
SENSICA
Manufacturer (Section D)
ADAPTEC MEDICAL DEVICES LLC 3014271001
260 james jackson avenue
cary NC 27513
MDR Report Key9538907
MDR Text Key178145872
Report Number1018233-2020-00002
Device Sequence Number1
Product Code EXS
Combination Product (y/n)N
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,other
Type of Report Initial,Followup
Report Date 02/21/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberSCCS1001
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/30/2020
Initial Date Manufacturer Received 12/10/2019
Initial Date FDA Received01/02/2020
Supplement Dates Manufacturer Received02/19/2020
Supplement Dates FDA Received02/21/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-