• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. TRAUMA IMPLANT; NAIL, FIXATION, BONE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. TRAUMA IMPLANT; NAIL, FIXATION, BONE Back to Search Results
Device Problem Fracture (1260)
Patient Problem Injury (2348)
Event Type  Injury  
Event Description
In a scientific publication, sellan, "short versus long intertan fixation for geriatric intertrochanteric hip fractures: a multicentre head-to-head comparison" it was reported that the patient that was using the short nail presented a periprosthetic fracture.It is unknown how this adverse event was treated.
 
Manufacturer Narrative
It was reported from a literature review that the patient using the short nail presented a periprosthetic fracture.It is unknown how this adverse event was treated.The study compared geriatric patients treated with either a short or a long intertan intramedullary nail fixation.The affected complaint device, used in treatment, was not returned for evaluation.Therefore a product analysis could not be performed.As device information was not made available, device history record and complaint history review cannot be completed.There is no information that would suggest the device failed to meet specifications.A relationship, if any, between the device and the reported incident could not be corroborated.A medical analysis noted that overall, most patients steadily improved their function and locomotive performance postoperatively with an intramedullary device for the treatment of a geriatric intertrochanteric hip fracture, independent of length of construct.It was concluded that the post-operative functional status of geriatric patients with intertrochanteric hip fractures was not influenced by the length of the device.Based on this investigation, the need for corrective action is not indicated.The potential root causes could include but are not limited to healing complications (nonunion, infection, etc.) that weakened bone, impeded fixation, or otherwise led to premature fracture.Without the return of the actual product involved and no patient medical records available, our investigation of this report is inconclusive.Based on this investigation, the need for corrective action is not indicated.No further investigation is warranted for this complaint; however we will continue to monitor for future complaints and investigate as necessary.Should the device or additional information be received, the complaint will be reopened.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRAUMA IMPLANT
Type of Device
NAIL, FIXATION, BONE
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks road
memphis TN 38116
MDR Report Key9544913
MDR Text Key173514020
Report Number1020279-2020-00061
Device Sequence Number1
Product Code JDS
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional,l
Type of Report Initial,Followup
Report Date 04/17/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 12/10/2019
Initial Date FDA Received01/03/2020
Supplement Dates Manufacturer Received04/16/2020
Supplement Dates FDA Received04/17/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Hospitalization;
-
-