• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW ORTHOPAEDICS LTD ACETLR CUP HAP 48MM W/ IMPTR; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/METAL, RESURFACING

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW ORTHOPAEDICS LTD ACETLR CUP HAP 48MM W/ IMPTR; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/METAL, RESURFACING Back to Search Results
Catalog Number 74120148
Device Problems Appropriate Term/Code Not Available (3191); Patient Device Interaction Problem (4001)
Patient Problems Pain (1994); Injury (2348)
Event Date 07/31/2019
Event Type  Injury  
Event Description
It was reported that a revision was performed due to persistent pain.
 
Manufacturer Narrative
It was reported that right hip revision surgery was performed.During the revision, the cup, head stem and sleeve were removed.As of today, the implanted devices, all of which were used in treatment, and additional information have been requested for this complaint but have not become available.A review of the complaint history for the acetabular cup, hemi head, stem and modular sleeve was performed using batch numbers in search of similar recurring reports for the products during their lifetimes.No other similar complaints were identified for the acetabular cup, hemi head and stem.Similar complaints have been identified for the modular sleeve.However, as the device is no longer sold, no action is to be taken.In the absence of the actual devices, the production records were reviewed for the devices reportedly involved in this incident.All the released devices involved met manufacturing specifications at the time of production.Review of the product ifu found adequate warnings and precautions in relation to the alleged failure modes.A risk management review was performed.No additional risks were identified as result of the reported event.No further actions are required at this time.The available medical documents were reviewed.The patient¿s reported persistent pain may be consistent with an adverse reaction to metal debris, but this cannot be confirmed based on the limited information provided.Without the supporting revision operative report, post-primary radiographic images, revision lab/pathology results, and/or the analysis of the explanted components, the root cause of the reported clinical symptoms cannot be confirmed and it cannot be concluded that the reported clinical reactions were associated with a mal-performance of the implant.The patient impact beyond the revision and associated post-op pain cannot be determined, and there is no report of the patient¿s current condition, or whether the reported clinical symptoms persist.Without return of the actual devices or further information we cannot further investigate or confirm the details supplied in this complaint, and our investigation remains inconclusive.If the products or additional information become available in the future, this case will be reopened.No preventative or corrective action has been initiated as a result of this investigation.
 
Manufacturer Narrative
New information: g4, d4.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACETLR CUP HAP 48MM W/ IMPTR
Type of Device
PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/METAL, RESURFACING
Manufacturer (Section D)
SMITH & NEPHEW ORTHOPAEDICS LTD
aurora house
spa park
leamington spa CV31 3HL
UK  CV31 3HL
MDR Report Key9548769
MDR Text Key173653896
Report Number3005975929-2020-00001
Device Sequence Number1
Product Code NXT
Combination Product (y/n)N
PMA/PMN Number
P040033
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type other
Type of Report Initial,Followup,Followup
Report Date 10/22/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Lay User/Patient
Device Expiration Date06/30/2013
Device Catalogue Number74120148
Device Lot Number089472
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 11/06/2019
Initial Date FDA Received01/03/2020
Supplement Dates Manufacturer Received07/02/2020
08/18/2020
Supplement Dates FDA Received07/07/2020
10/22/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Treatment
71356103 ANTH HO PO SZ 3 09LM11606; 74122542 HEMI HEAD 42MM 08FW17356; 74222200 MOD SLEEVE P 0MM 12/14 09BW22223
Patient Outcome(s) Hospitalization; Required Intervention;
-
-