Model Number 2426-0007 |
Device Problems
Fluid/Blood Leak (1250); Stretched (1601); Material Deformation (2976)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 12/13/2019 |
Event Type
malfunction
|
Manufacturer Narrative
|
Patient information requested but not provided.The affected product has been received and the investigation is pending.A follow up report will be submitted once the failure investigation has been completed.
|
|
Event Description
|
It was reported that there was a large bubble in the tubing close to the valve and it sprayed the staff as they were trying to connect.There was no patient injury with this event.
|
|
Manufacturer Narrative
|
The customer¿s report that there was a large bubble in the tubing was confirmed by visual inspection.The customer also reported that the tubing sprayed (leaked) but limited functional testing did not observe any leaks.The set was visually inspected for kinks, incomplete bonding engagements, holes/tears in the tubing or damages to the components.At the top of the silicone pump segment tubing near the upper fitment, a balloon was observed.Functional testing was performed and the set reprimed successfully via gravity with no signs of leaks or anomalies.Previous failure investigations have found that ballooning can occur when an iv push medication or flush is executed below the pump without first clamping the tubing above the injection port.This action was found to result in excessive pressure within the silicone segment thus causing the silicone segment to balloon.The root cause for the source of the excessive pressure is unknown.The customer¿s report that the tubing sprayed (leaked) was not confirmed.The root cause was not identified.
|
|
Event Description
|
It was reported that there was a large "bubble" in the tubing close to the valve and it sprayed the staff as they were trying to connect.It was further confirmed, that there was no patient harm as a result of this event.
|
|
Search Alerts/Recalls
|