This device was thoroughly inspected and analyzed.Review of device memory confirmed that the device had reached elective replacement indicator (eri) status due to long charge times, during a period of magnet application for two and a half hours.A manual capacitor reform was performed, and the charge time was 9.4 seconds.A visual inspection of the device header and case noted no anomalies.Pin gauge testing, designed to verify proper port dimensions, was completed.Each port measured as expected.The device was then exposed to simulated heart load conditions, and the defibrillation, pacing, and sensing functions were tested.The device operated appropriately with no interruptions in therapy output at the returned programmed settings.A series of electrical tests was also performed, and again, normal device function was observed.
|