Avanos medical inc.Received a single report that referenced 12 different incidences, which were associated with separate units, involving 12 different patients.This is the sixth of twelve reports.Refer to 3011270181-2020-00001 for the first report.Refer to 3011270181-2020-00002 for the second report.Refer to 3011270181-2020-00003 for the third report.Refer to 3011270181-2020-00004 for the fourth report.Refer to 3011270181-2020-00005 for the fifth report.Refer to 3011270181-2020-00007 for the seventh report.Refer to 3011270181-2020-00008 for the eighth report.Refer to 3011270181-2020-00009 for the ninth report.Refer to 3011270181-2020-00010 for the tenth report.Refer to 3011270181-2020-00011 for the eleventh report.Refer to 3011270181-2020-00012 for the twelfth report.It was reported in "undetected cortrak tube misplacements in the (b)(6), (b)(6) 2010: an audit of trace interpretation" (b)(6) that sometime between 2010 and 2017, a nasogastric tube was placed in the patient's lung.Placement was not confirmed by x-ray, and the patient received feed.The patient died.No further information was provided.
|