• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW ORTHOPAEDICS LTD FEMORAL HEAD 52MM; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/METAL, RESURFACING

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW ORTHOPAEDICS LTD FEMORAL HEAD 52MM; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/METAL, RESURFACING Back to Search Results
Catalog Number 74123152
Device Problems Appropriate Term/Code Not Available (3191); Patient Device Interaction Problem (4001)
Patient Problems Pain (1994); Toxicity (2333); Injury (2348); Ambulation Difficulties (2544)
Event Date 02/28/2017
Event Type  Injury  
Event Description
It was reported a left hip revision surgery due to pain, limited mobility, metallosis and elevated metal ion levels.
 
Manufacturer Narrative
It was reported that left hip revision surgery was performed.As of today, the implanted devices, all of which were used in treatment, and additional information have been requested for this complaint but have not become available.A review of the complaint history for the bhr cup and bhr head was performed using batch numbers in search of similar recurring reports for the products during their lifetimes.No other similar complaints were identified for the head or cup.In the absence of the actual devices, the production records were reviewed for the devices reportedly involved in this incident.All the released devices involved met manufacturing specifications at the time of production.Review of the product ifu found adequate warnings and precautions in relation to the alleged failure modes.A risk management review was performed.No additional risks were identified as result of the reported event.The available medical documents were reviewed.The revision report did not note findings related to metal related pathology.The root cause of the reported pain and limited mobility cannot be confirmed and it cannot be concluded that the reported findings are related to the s+n product based on the information provided.The patient impact beyond the pain, revision, and expected transient post-op convalescence period cannot be determined.Without return of the actual devices or further information we cannot further investigate or confirm the details supplied in this complaint, and our investigation remains inconclusive.If the products or additional information become available in the future, this case will be reopened.No preventative or corrective action has been initiated as a result of this investigation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FEMORAL HEAD 52MM
Type of Device
PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/METAL, RESURFACING
Manufacturer (Section D)
SMITH & NEPHEW ORTHOPAEDICS LTD
aurora house
spa park
leamington spa CV31 3HL
UK  CV31 3HL
MDR Report Key9553354
MDR Text Key173774774
Report Number3005975929-2020-00004
Device Sequence Number1
Product Code NXT
Combination Product (y/n)N
PMA/PMN Number
P040033
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type other
Type of Report Initial,Followup,Followup
Report Date 10/22/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Lay User/Patient
Device Expiration Date03/30/2015
Device Catalogue Number74123152
Device Lot Number10CW25932
Initial Date Manufacturer Received 12/30/2019
Initial Date FDA Received01/06/2020
Supplement Dates Manufacturer Received06/30/2020
08/18/2020
Supplement Dates FDA Received07/04/2020
10/22/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Treatment
74122158 ACET CUP HAP SIZE 52/58 10AW25562
Patient Outcome(s) Hospitalization; Required Intervention;
-
-