Blank fields on this form indicate the information is unchanged, unknown or unavailable.Investigation ¿ evaluation.It was reported that an unknown ultrathane catheter experienced hub separation.The patient was taken to the er and the device was replaced.Cook became aware of this event upon being notified by hospice & palliative care charlotte region.The patient reportedly experienced no additional adverse effects as a result of this incident, no additional harm was reported.A review of the drawing, instructions for use (ifu), manufacturing instructions, quality control, and specifications of the device were conducted during the investigation the complaint device was not returned so a physical examination could not be performed.Additionally, a document-based investigation evaluation was performed.Sufficient inspection activities are in place to identify this failure mode prior to distribution.Cook could not review the device history record (dhr) due to lack of lot information from the user facility.Review of the sales records to the user facility over the past three years could not sufficiently narrow down the lot number.Since potential nonconformances or complaints from the complaint device¿s lot could not be confirmed, there is no evidence that nonconforming product exists in house or in the field.A capa was previously opened to address this failure and concluded that the main root cause was manufacturing related.Corrective actions including implementation of a gap gauge and retraining were performed.However, due to an unknown lot number, it is unknown whether or not the complaint device lot was manufactured prior to implementation of the corrective actions.Based on the information provided, no returned product, and results of the investigation, it was concluded that a manufacturing and quality control deficiency potentially contributed to this incident.The appropriate personnel have been notified.Per the quality engineering risk assessment no further action is required.Cook will continue to monitor for similar complaints.This report is required by the fda under 21 cfr part 803 and is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement contained herein is intended to be an admission that any cook device is defective or malfunctioned, that a death or serious injury occurred, nor that any cook device caused, contributed to, or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
|