Model Number 311.01 |
Device Problem
Device-Device Incompatibility (2919)
|
Patient Problems
No Consequences Or Impact To Patient (2199); No Patient Involvement (2645)
|
Event Type
malfunction
|
Manufacturer Narrative
|
Product complaint # (b)(4).Complainant part is not expected to be returned for manufacturer review/investigation.Without a lot number, the device history records review could not be completed as no product was received.Investigation summary product was not returned.Based on the information available, it has been determined that no corrective and preventive action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
|
Event Description
|
It was reported that on an unknown date, the handle with mini quick coupling was stuck.It is unknown if there was surgical delay.Procedure outcome was unknown.There was no patient consequence.This complaint involves one (1) device.This report is 1 of 1 for (b)(4).
|
|
Manufacturer Narrative
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.H3, h4, h6: a device history record (dhr) review was conducted: part # 311.01.Synthes lot # 7684810.Supplier lot # na.Release to warehouse date: 14 may 2014.Manufactured by synthes jennersville.No ncr's were generated during production.Review of the device history record(s) showed that there were no issues during the manufacture of this product, and any sub-components, which would contribute to this complaint condition.H3, h6: a product investigation was conducted.Visual inspection: the handle with mini quick coupling (p/n: 311.01, lot number: 7684810) was received at us cq.Upon visual inspection, the device is bent and deformed.No mating devices were returned to perform a functional test, but this complaint is being confirmed since the device condition can result in the complaint condition.Dimensional inspection: no dimensional inspection was performed due to post-manufacturing damage.Document/specification review: relevant drawings were reviewed.No design issues or discrepancies were identified.Service and repair evaluation: the customer reported the handle with mini quick coupling was stuck.The repair technician reported the device's tip was bent.Bent condition is the reason for repair.The item is not repairable per the inspection sheet.The cause of the issue is unknown.The item will be forwarded to customer quality.The evaluation was confirmed.Investigation conclusion: this complaint is confirmed as the device tip is bent.No definitive root cause could be determined based on the provided information.No new, unique or different patient harms were identified as a result of this evaluation.There was no indication that a design or manufacturing issue contributed to the complaint.No design issues were observed during the document/specification review.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post market safety surveillance activities.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
|
Manufacturer Narrative
|
Product complaint #: (b)(4).Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.G510(k) added to for reporting.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
|