B.BRAUN SURGICAL SA HISTOACRYL BLUE 0.5ML; TISSUE ADHESIVES Back to Search Results
Model Number 1050044
Device Problem Fluid/Blood Leak (1250)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Type  malfunction  
Manufacturer Narrative
Pma/510k: reported device not marketed in the u.S., however, similar devices or devices that share components, raw materials, process methods or other technological characteristics are registered within the u.S.K111959.If additional information becomes available, a follow up report will be submitted.
 
Event Description
It was reported ampoule leakage occurred.The reporter indicated that when the package was opened, it was found that the glue inside the package is missing and the leaking liquid had solidified.The event occurred before use with no patient harm.
 
Manufacturer Narrative
Investigation: no samples or picture showing the defect have been received.Reviewed the batch manufacturing record, this product had a normal process and the results during the process fulfill b.Braun surgical requirements.Final conclusion: without samples we are not in position of studying if the affected product does not fulfil the specifications.In consequence, a proper analysis cannot be done and the case is not confirmed due to lack of evidence.Nevertheless, we take note of this incidence and if any sample is received in the future, we will re-open the case and analyze it.Please note that when no samples are received our analysis is very limited.We regret any inconvenience this issue may have caused and thank you for your collaboration.Actions on product distributed of this reference/batch: based on the conclusion derived from investigation, it is not required to make actions in distributed product.No corrective/preventive actions needed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HISTOACRYL BLUE 0.5ML
Type of Device
TISSUE ADHESIVES
Manufacturer (Section D)
B.BRAUN SURGICAL SA
121 carretera de terrassa
rubi, 08191
SP  08191
MDR Report Key9555585
MDR Text Key173939462
Report Number3003639970-2019-00967
Device Sequence Number1
Product Code MPN
Combination Product (y/n)N
PMA/PMN Number
SEE H10
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 03/23/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/06/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number1050044
Device Catalogue Number1050044
Device Lot Number218406N2
Was Device Available for Evaluation? No
Date Manufacturer Received03/20/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-