The reported event was confirmed.Visual evaluation of the sample noted one bulb syringe received in opened packaging.There were 5 specks of foreign material found in the syringe on and under the bulb measuring 0.80 sq.Mm, 0.60 sq.Mm, and 3 of them measuring 0.20 sq.Mm.This did not meet the specification "loose foreign matter or embedded shall not exceed an aggregate total of 0.6mm² or 1/16¿.Although the reported event was confirmed, the root cause could not be determined.A potential root cause for this failure could be defective / contaminated components from supplier.The device history record was reviewed and found nothing that could have caused or contributed to the reported event.A labeling review was not performed because labelling could not have prevented the reported failure.H11:section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|