• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ARROW INTERNATIONAL INC. EPIDURAL CATHETERIZATION SET; ANESTHESIA CONDUCTION KIT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ARROW INTERNATIONAL INC. EPIDURAL CATHETERIZATION SET; ANESTHESIA CONDUCTION KIT Back to Search Results
Model Number IPN046311
Device Problem Fluid/Blood Leak (1250)
Patient Problem No Information (3190)
Event Date 01/06/2020
Event Type  malfunction  
Manufacturer Narrative
Qn#: (b)(4).The device has not been returned for investigation.Teleflex will continue to monitor and trend related events.
 
Event Description
The report states: we are having leak issues with the syringes from epidural catheter kit flextip plus (jc-05400-b).The plunger gets loose easily.Several incidents reported by nurses involving lot 71f19k1979.
 
Manufacturer Narrative
(b)(4).A device history record review was performed on the lor syringe with no relevant findings.The customer reported the lor syringe leaked.The customer returned one 10ml plastic lor syringe and lidstock (reference files (b)(4)).The syringe was visually examined.Visual examination of the syringe revealed that the syringe appears typical with no observed defects or anomalies.The returned sample was returned to the supplier (preox) for function testing.According to the supplier, "no leakage found by normal handling, leakage if plunger moved to the side".A design history review was performed for part # kz-05501-002 as a part of this complaint investigation.Per (b)(4) (released 03-dec-2018), supplier (preox) made the following changes: kz-05501-002 luer plastic lor syringe: a.Changed plunger material from profax 535 to profax 531 b.Changed to new plunger tool c.Changed to new mold for blue stopper d.Changed molding location for the plunger and the blue stopper as follows.Plunger: from fleimaplastic in germany to gpe, germany, blue stopper: from et, germany to psilkon, germany.These effective changes did impact product design and material.It should be noted, the returned lor syringe was from the new design.The reported complaint of the lor syringe leaking was confirmed based on the sample received.The returned lor syringe was returned to the supplier (preox) for functional testing where according to the supplier, "no leakage found by normal handling, leakage if plunger moved to the side".A device history record review was performed on the lor syringe with no evidence to suggest a manufacturing related issue.However, based on the functional testing by the supplier, the potential root cause of this issue is supplier related.A scar has been initiated to further investigate this issue.
 
Event Description
The report states: we are having leak issues with the syringes from epidural catheter kit flextip plus (jc-05400-b).The plunger gets loose easily.Several incidents reported by nurses involving lot 71f19k1979.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EPIDURAL CATHETERIZATION SET
Type of Device
ANESTHESIA CONDUCTION KIT
Manufacturer (Section D)
ARROW INTERNATIONAL INC.
reading PA
MDR Report Key9559006
MDR Text Key179718858
Report Number3006425876-2020-00063
Device Sequence Number1
Product Code CAZ
Combination Product (y/n)N
PMA/PMN Number
K103658
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional,u
Type of Report Initial,Followup
Report Date 01/06/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date10/10/2021
Device Model NumberIPN046311
Device Catalogue NumberJC-05400-B
Device Lot Number71F19K1979
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/28/2020
Initial Date Manufacturer Received 01/06/2020
Initial Date FDA Received01/07/2020
Supplement Dates Manufacturer Received02/20/2020
Supplement Dates FDA Received02/25/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-