Device returned to manufacturer: the comet was returned without the handle and analysis was completed.The tip, device shaft and sensor port were examined for damage or any irregularities.The investigation of the shaft of the device showed no damage.The tip showed a bend.It was noticed that there was some peeling of the coating located 81cm from the tip and continued along the shaft to 131cm.Inspection of the remainder of the device, apart from the observed damage, revealed no other damage or irregularities.Because there was no evidence of any product quality deficiencies, it was considered likely that the peeled coating and the tip damage was attributable to handling issues; therefore, the conclusion code unintended use error caused or contributed to event.
|