Model Number 95214
Device Problems Fluid/Blood Leak (1250); Insufficient Flow or Under Infusion (2182); Infusion or Flow Problem (2964)
Patient Problem Low Oxygen Saturation (2477)
Event Date 11/28/2019
Event Type  Injury  
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
Medtronic received information that during heart surgery it was reported that the oxygenation capacity gradually decreased with use of an affinity nt oxygenator.The carbon dioxide removal capacity decreased and plasma leaked from the lower end of the oxygenator.The customer constantly increased the oxygen concentration and increased the air flow to compensate for gas exchange, and a blood transfusion was administered to compensate for blood oxygen and coagulation function.Patient outcome was described as alive, no injury.
 
Manufacturer Narrative
B5.The circuit was primed with blood for 5 hours.Heparin sodium injection, compound electrolyte injection, injection methylprednisolone sodium succinate.No visible air in the circuit flow rates 0.8-5.6l,pressures 80-185mmhg medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Manufacturer Narrative
D4.Serial number updated medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AFFINITY NT I-CVR/HFO 541
Type of Device
OXYGENATOR, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
PERFUSION SYSTEMS
7611 northland dr
brooklyn park MN 55428
MDR Report Key9563832
MDR Text Key174178961
Report Number2184009-2020-00004
Device Sequence Number1
Product Code DTZ
Combination Product (y/n)N
PMA/PMN Number
K191029
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup,Followup
Report Date 03/11/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date04/24/2020
Device Model Number95214
Device Catalogue Number95214
Device Lot Number13256543
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 12/10/2019
Initial Date FDA Received01/08/2020
Supplement Dates Manufacturer Received01/10/2020
02/13/2020
Supplement Dates FDA Received02/05/2020
03/11/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age49 YR
-
-