ABBOTT DIABETES CARE INC FREESTYLE LIBRE 14 DAY; FLASH GLUCOSE MONITORING SYSTEM Back to Search Results
Model Number 71940-01
Device Problem Appropriate Term/Code Not Available (3191)
Patient Problems Abscess (1690); Unspecified Infection (1930)
Event Date 12/17/2019
Event Type  Injury  
Manufacturer Narrative
The reported product is not expected to be returned as the device was reportedly discarded.A follow-up report will be submitted if product is returned or additional information is obtained the date of event is unknown.The date entered is the date abbott diabetes care became aware of the event.The device mfg date is unknown.The date entered is the date abbott diabetes care became aware of the event.All pertinent information available to abbott diabetes care has been submitted.
 
Event Description
Customer reported experiencing a skin reaction while wearing an adc freestyle libre sensor and experienced an abscess at the sensor insertion site.Customer had contact with a healthcare professional and received unspecified antibiotics for treatment.There was no report of death or permanent injury associated with this event.
 
Manufacturer Narrative
No product has been returned and a valid serial number has not been provided.Extended investigation has been performed for the reported complaint and there was no indication that the product did not meet specification.The reported complaint is related to skin irritation or an allergic reaction to the patch adhesive of the freestyle libre sensor.Dose audit reports were reviewed and demonstrated the continued effectiveness of the established sterilization process for libre sensor products.Environmental monitoring reports were reviewed, including bioburden and endotoxin testing, and demonstrated that all monitoring processes continue to meet adc minimum requirements for product quality.Clinical data was reviewed and confirmed that freestyle libre sensor continue to be safe, effective, and perform as intended in the field.A tripped trend review was completed for the reported complaint and fs libre sensors, and there were no adverse trends that indicate any potential product related issues.If the product is returned, the case will be re-opened and a physical investigation will be performed.
 
Event Description
Customer reported experiencing a skin reaction while wearing an adc freestyle libre sensor and experienced an abscess at the sensor insertion site.Customer had contact with a healthcare professional and received unspecified antibiotics for treatment.There was no report of death or permanent injury associated with this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FREESTYLE LIBRE 14 DAY
Type of Device
FLASH GLUCOSE MONITORING SYSTEM
Manufacturer (Section D)
ABBOTT DIABETES CARE INC
1360 south loop road
alameda CA 94502 7001
Manufacturer Contact
tara williamson
1360 south loop road
alameda, CA 94502-7001
5108644472
MDR Report Key9566183
MDR Text Key174397952
Report Number2954323-2020-00290
Device Sequence Number1
Product Code PZE
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P160030
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 02/26/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Lay User/Patient
Device Model Number71940-01
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 12/17/2019
Initial Date FDA Received01/08/2020
Supplement Dates Manufacturer Received02/05/2020
Supplement Dates FDA Received02/26/2020
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured12/17/2019
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Weight84
-
-