The distributor in (b)(4) rejected 130344, single function abc pencil, due to an "insufficient heat seal".In this instance, there was no patient involvement as the packaging anomaly was discovered during incoming inspection prior to distribution to an end-user.This report is being raised based on device malfunction with potential for injury upon reoccurrence.
|
Received one 130344 in original packaging.Lot number was verified.Performed a visual inspection, there were no obvious signs of abnormalities or defects.Performed a functional inspection, the devices were dye leak tested, which indicated that the packaging did not have an insufficient heat seal.The manufacturing documents from the device history record have been reviewed with special attention to the manufacturing and inspection of the product.The product released for distribution were found to have met all specifications prior to shipment.This is the only complaint for this lot number and failure mode within the past two years.A two-year review of complaint history revealed there has been 7 complaints regarding 8 devices for this device family and failure mode.During the same time frame 207,719 devices have been manufactured and shipped worldwide.Should all the complaint devices have been found confirmed, the rate of failure would be.00004 per the instructions for use, the user is advised the following; - these devices should be inspected before use.Visually examine the devices for obvious physical damage and do not if damage is found: cracked, broken or otherwise distorted plastic parts; broken or significantly bent handle, shaft or connector contacts.This issue will continue to be monitored through the complaint system to assure patient safety.
|