TERUMO CARDIOVASCULAR SYSTEMS CORPORATION ADVANCED PERFUSION SYSTEM 1; CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS- ADVANCED PERFUSION SYSTEM 1 Back to Search Results
Model Number 816571
Device Problem Noise, Audible (3273)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 11/11/2019
Event Type  malfunction  
Manufacturer Narrative
Per data log analysis, there are no indications of a problem in the pump log.Normally a noisy pump will not cause any unusual log entries.During laboratory analysis, the product surveillance technician (pst) observed a loud rattling noise inside the pump when tilted.The pump was opened and it was observed that the pump gut was sheared off at the stop nut and washer.If additional information becomes available on this complaint that would alter the facts and/or conclusion, a supplemental report will be filed accordingly.
 
Event Description
It was reported that during priming of the device for a cardiopulmonary bypass (cpb) procedure, the roller pump had a grinding noise.As a result, an alternate device was employed.The surgical procedure was completed successfully.There was no delay, no blood loss, nor adverse consequences to the patient.
 
Manufacturer Narrative
The reported complaint was confirmed.If additional information becomes available on this complaint that would alter the facts and/or conclusion, a supplemental report will be filed accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ADVANCED PERFUSION SYSTEM 1
Type of Device
CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS- ADVANCED PERFUSION SYSTEM 1
Manufacturer (Section D)
TERUMO CARDIOVASCULAR SYSTEMS CORPORATION
6200 jackson road
ann arbor MI 48103
MDR Report Key9572162
MDR Text Key198598752
Report Number1828100-2020-00011
Device Sequence Number1
Product Code DTQ
UDI-Device Identifier00886799001370
UDI-Public(01)00886799001370(11)190225
Combination Product (y/n)N
PMA/PMN Number
K172220
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 01/28/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number816571
Device Catalogue Number816571
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/25/2019
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 01/02/2020
Initial Date FDA Received01/09/2020
Supplement Dates Manufacturer Received01/14/2020
Supplement Dates FDA Received01/28/2020
Is This a Reprocessed and Reused Single-Use Device? No
Removal/Correction Number1828100-12/09/19-004-C
Patient Sequence Number1
-
-