TERUMO CARDIOVASCULAR SYSTEMS CORPORATION ADVANCED PERFUSION SYSTEM 1; CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS- ADVANCED PERFUSION SYSTEM 1
|
Back to Search Results |
|
Model Number 816571 |
Device Problem
Noise, Audible (3273)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 11/11/2019 |
Event Type
malfunction
|
Manufacturer Narrative
|
Per data log analysis, there are no indications of a problem in the pump log.Normally a noisy pump will not cause any unusual log entries.During laboratory analysis, the product surveillance technician (pst) observed a loud rattling noise inside the pump when tilted.The pump was opened and it was observed that the pump gut was sheared off at the stop nut and washer.If additional information becomes available on this complaint that would alter the facts and/or conclusion, a supplemental report will be filed accordingly.
|
|
Event Description
|
It was reported that during priming of the device for a cardiopulmonary bypass (cpb) procedure, the roller pump had a grinding noise.As a result, an alternate device was employed.The surgical procedure was completed successfully.There was no delay, no blood loss, nor adverse consequences to the patient.
|
|
Manufacturer Narrative
|
The reported complaint was confirmed.If additional information becomes available on this complaint that would alter the facts and/or conclusion, a supplemental report will be filed accordingly.
|
|