It was reported that when this dpt (disposable pressure transducer) was connected to a right femoral line of a (b)(6) female patient, the systolic and diastolic pressure values did not match with the pressure values obtained with a non-invasive bp cuff.The sensor was zeroed and calibrated, and the line was flushed.However, the values shown on the monitor were too high and were not consistent with the patient's condition.The line was flushed several more times, but the same problem was still present, so they requested another transducer with which the values were different and more accurate to the patient's condition.The device was available for evaluation.
|
One single dpt kit was returned for examination.The reported event of "pressure values and its waveform were too high" was not confirmed.The dpt sensor zeroed and sensed pressure accurately on a pressure monitor.The pressure did not show any drift during output drift testing and met specification.Electrical testing showed that both input and output impedance of the dpt sensor were within specifications.Zero-offset also met specification.No leakage or occlusion was detected from the dpt kit during pressure test.No visible damage was observed from the dpt kit.Dry blood was noted at the stand-alone stopcock and distal connector of the pressure line.The reported event could not be confirmed or replicated during the analysis, as the device responded appropriately during functional testing.There was no evidence of a manufacturing nonconformance.No further actions will be taken at this time.Poor dynamic response can be caused by air bubbles, clotting, and excessive lengths of tubing, excessively compliant pressure tubing, small bore tubing, loose connections, or leaks.The assembly may be tested for dynamic response by observing the pressure waveform on an oscilloscope or monitor.Bedside determination of the dynamic response of the catheter, monitor, kit and transducer system is done after the system is flushed, attached to the patient, zeroed and calibrated.A square-wave test may be performed by pulling the snap tab device and releasing quickly.Pressure readings can change quickly and dramatically because of loss of proper calibration, loose connection, or air in the system.Pressure readings should correlate with the patient¿s clinical manifestations.It is not known if user or procedural factors may have contributed to the stated event.In this event, there was no patient compromise noted.Complaint histories for all reported events are reviewed against trending control limits on a monthly basis, and any excursions above the control limits are assessed and documented as part of this monthly review.
|