• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. 3.5MM S/R HEXDRIVER SH 178MM; SCREWDRIVER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. 3.5MM S/R HEXDRIVER SH 178MM; SCREWDRIVER Back to Search Results
Model Number 71173487
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 01/07/2020
Event Type  malfunction  
Event Description
It was reported that during surgery, tip of driver broke off inside the patient.All pieces recovered.S&n backup device available.No delay.No impact or injury to patient reported.
 
Manufacturer Narrative
The associated device, used in treatment, was returned and evaluated.A visual inspection of the returned device confirmed the stated failure mode.The hex tip on the device fractured off, rendering the device inoperable.The tip was not returned.The device was manufactured in 2011 and shows signs of extensive wear / usage.A review of complaint history did not reveal additional complaints for the listed batch.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.This device is a reusable instrument that can be exposed to numerous surgeries.Damage from prolonged use, misuse or rough handling are likely probable causes of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
3.5MM S/R HEXDRIVER SH 178MM
Type of Device
SCREWDRIVER
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks road
memphis TN 38116
MDR Report Key9580481
MDR Text Key174684402
Report Number1020279-2020-00232
Device Sequence Number1
Product Code HXX
UDI-Device Identifier03596010629753
UDI-Public03596010629753
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Type of Report Initial,Followup
Report Date 02/06/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number71173487
Device Catalogue Number71173487
Device Lot Number11LM01342
Initial Date Manufacturer Received 01/07/2020
Initial Date FDA Received01/13/2020
Supplement Dates Manufacturer Received02/03/2020
Supplement Dates FDA Received02/06/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-