Model Number 214600
Device Problem Difficult to Open or Close (2921)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 01/02/2020
Event Type  malfunction  
Manufacturer Narrative
Product complaint # (b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.(b)(4).
 
Event Description
It was reported by the sales rep via phone that during a rotator cuff repair procedure both their penetrating grasper straight and penetrating grasper 35 degree up will not open or close.The procedure was completed with another angle device with no patient harm or surgical delay to the case.The devices will be returning for evaluation.
 
Manufacturer Narrative
Product complaint # (b)(4).Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Investigation summary: according to the information provided, it was reported that during a rotator cuff repair procedure both their penetrating grasper straight and penetrating grasper 35 degree up will not open or close.The complaint device was received and evaluated.Visual observation of the arthro grasper pen str it was found a slighty bent on the wand of the device but reveal worn, used but in expected condition.The device was functional tested, and can be confirming this complaint as the arthro grasper pen str was the jaws not open/close.The jaws do not operate as intended.At the moment to open and close the jaws with the trigger, the upper jaw is loose and need to put in the position as a manually form.The possible root cause of this type of failure can be attributed to frequent exerting too much pressure in operating the device, it was related possibly heavily used.Given that no lot number was provided, a manufacturing record evaluation (mre) review cannot be performed.If the lot number becomes available, the mre review will be performed.At this point in time, no corrective action is required, and no further action is warranted.However, depuy synthes mitek will continue to track any related complaints within this device family as a means of monitoring the extent with which this complaint is observed in the field.Udi: (b)(4).
 
Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.H10 additional narrative: d10, h3, h6: the actual device has been returned and is currently pending evaluation.Once reliability engineering evaluates the device, a supplemental medwatch report will be sent accordingly.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Type of Device
SUTURE/NEEDLE PASSER, REUSABLE
Manufacturer (Section D)
DEPUY MITEK LLC US
325 paramount drive
raynham MA 02767
MDR Report Key9587209
MDR Text Key202749722
Report Number1221934-2020-00176
Device Sequence Number1
Product Code NBH
UDI-Device Identifier10886705003839
UDI-Public10886705003839
Combination Product (y/n)N
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,other
Type of Report Initial,Followup,Followup
Report Date 01/02/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number214600
Device Catalogue Number214600
Device Lot Number201
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/14/2020
Initial Date Manufacturer Received 01/02/2020
Initial Date FDA Received01/14/2020
Supplement Dates Manufacturer Received02/14/2020
02/27/2020
Supplement Dates FDA Received02/18/2020
02/28/2020
Patient Sequence Number1
-
-