Model Number 16-02-80 |
Device Problem
Infusion or Flow Problem (2964)
|
Patient Problem
No Patient Involvement (2645)
|
Event Date 12/12/2019 |
Event Type
malfunction
|
Manufacturer Narrative
|
There was no patient involvement.The heater-cooler 16-02-80 is not distributed in the usa and it is similar to heater-cooler 16-02-85, which is distributed in the usa (510(k) number: k052601).Livanova (b)(4) manufactures the heater-cooler 16-02-80.The incident occurred in (b)(6).A livanova field service representative was dispatched to the facility to investigate the device.He replaced the patient pump 1 and the distribution board and the issue could be solved.Functional verification testing was completed without further issues and the unit was returned to service.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.
|
|
Event Description
|
Livanova (b)(4) received a report that a heater-cooler system 3t displayed an error code on the patient side during maintenance.There was no patient involvement.
|
|
Manufacturer Narrative
|
H.10: through follow-up communication livanova learned that the distribution board was most likely damaged by the defective patient pump 1.Based on all the information available, it is reasonable to assume that a defective electronics of the patient pump is the root cause of the reported event.
|
|