• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW ORTHOPAEDICS LTD ACETABULAR CUP HAP SIZE 44/50; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/METAL, RESURFACING

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW ORTHOPAEDICS LTD ACETABULAR CUP HAP SIZE 44/50; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/METAL, RESURFACING Back to Search Results
Model Number 74122150
Device Problems Device Dislodged or Dislocated (2923); Adverse Event Without Identified Device or Use Problem (2993)
Patient Problems Failure of Implant (1924); Test Result (2695)
Event Date 01/10/2020
Event Type  Injury  
Event Description
It was reported that a conversion bhr to tha was performed due to elevated test results (co 8 cr 9.2), cup looks to slipped and now retroverted.
 
Manufacturer Narrative
It was reported that hip revision surgery was performed.As of today, the implanted devices, all of which were used in treatment, and additional information have been requested for this complaint but has not become available.Without definitive lot numbers a complete complaint history review cannot be performed for the devices involved.A review of the complaint history review was performed using the part number for a cup and head in search of complaints involving elevated tests results throughout the lifetime of the product.As no device batch numbers were provided for investigation, a manufacturing record review, ifu review and risk management review could not be performed.If more information is received, this investigation will be reopened.No medical records have been received on this complaint.It was reported that a revision was performed due to elevated test results.However the device will not be returned for analysis.Without supporting medical records/documentation, the reported event cannot be assessed and a thorough medical assessment cannot be performed.Without return of the actual devices or further information we cannot further investigate or confirm the details supplied in this complaint, and our investigation remains inconclusive.If the products or additional information become available in the future, this case will be reopened.No preventative or corrective action has been initiated as a result of this investigation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACETABULAR CUP HAP SIZE 44/50
Type of Device
PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/METAL, RESURFACING
Manufacturer (Section D)
SMITH & NEPHEW ORTHOPAEDICS LTD
aurora house
spa park
leamington spa CV31 3HL
UK  CV31 3HL
MDR Report Key9592691
MDR Text Key175099315
Report Number3005975929-2020-00018
Device Sequence Number1
Product Code NXT
UDI-Device Identifier03596010552273
UDI-Public03596010552273
Combination Product (y/n)N
PMA/PMN Number
P040033
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 02/25/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/15/2020
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator No Information
Device Model Number74122150
Device Catalogue Number74122150
Was Device Available for Evaluation? Yes
Date Manufacturer Received01/10/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Treatment
FEMORAL HEAD 74123144/ UNKNOWN; FEMORAL HEAD 74123144/ UNKNOWN
Patient Outcome(s) Hospitalization; Required Intervention;
-
-