The reported event was unconfirmed as the problem could not be reproduced.Evaluation found no leakage was observed even when pressure was applied on the sac.The sample had undergone a 7 day leak test and no leakage was observed.There was also no leakage from the valve.The catheter was dissected and no conditions were found that could have contributed to the reported event.The device history record was reviewed and found nothing that could have caused or contributed to the reported event.The instructions for use were found adequate and state the following: "sterile: unless package is opened or damaged.Do not use if package is opened or damaged.Warning: on catheter, do not use ointments or lubricants having petrolatum base.They will damage latex and may cause the balloon to burst.Caution: do not aspirate urine through drainage funnel wall.Storage: store catheters at room temperature away from direct exposure to light, preferably in the original box.Visually inspect the product for any imperfections or surface deterioration prior to use.Valve type: use luer slip syringe.Do not use needle.Recommended inflation capacities: - 3cc balloon: use 5cc sterile water - 5cc balloon: use 10cc sterile water - 30cc balloon: use 35cc sterile water" h11:section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|