The reported event was confirmed as use-related.The customer stated that the purewick was used for 28 days.According to the ifu, "replace the purewicktm female external catheter every 8-12 hours or when soiled with feces or blood." a potential root cause could be due to the "user unaware of time limitation or forgets the patient is using the device." as no sample was returned it is unknown if the device met specifications; however, based on the ifu there appears to be to use-related relationship between the reported event and the device.Based on the event ,it appears that the device was used for treatment.The lot number is unknown therefore the device history record could not be reviewed.The instructions for use were found adequate and state the following: ¿setup: 1.Connect the canister to wall suction and set to a minimum of 40mmhg continuous suction.Always use the minimum amount of suction necessary.If using the drydoctm vacuum station, connect the canister to the unit and turn the unit on.Please consult the drydoctm vacuum station user guide for further information.2.Using standard suction tubing, connect the purewicktm female external catheter to the collection canister.Peri-care and placement: 3.Perform perineal care and assess skin integrity (document per hospital protocol).Separate legs, gluteus muscles, and labia.Palpate pubic bone as anatomical marker.4.With soft gauze side facing patient, align distal end of the purewicktm female external catheter at gluteal cleft.Gently tuck soft gauze side between separated gluteus and labia.Ensure that the top of the gauze is aligned with the pubic bone.Slowly place legs back together once the purewicktm female external catheter is positioned.Note: patient can be positioned on back, side lying, frog legged, or lying on back with knees bent and thighs apart (lithotomy position) prior to device placement.Maintenance: 6.Replace the purewicktm female external catheter at least every 8-12 hours or if soiled with feces or blood.Always assess skin for compromise and perform perineal care prior to placement of a new purewicktm female external catheter.Removal: 5.To remove the purewicktm female external catheter, fully separate the legs, gluteus, and labia.To avoid potential skin injury upon removal, gently pull the purewicktm female external catheter directly outward.Ensure suction is maintained while removing the purewicktm female external catheter.After use, this product may be a potential biohazard.Dispose of in accordance with applicable local, state and federal laws and regulations." h11:section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|