Catalog Number 00882100100
Device Problems Component Missing (2306); Output below Specifications (3004)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 09/13/2019
Event Type  malfunction  
Manufacturer Narrative
This event has been recorded by zimmer biomet under (b)(4).Udi: no: (b)(4).Telephone number: (b)(6).On (b)(6) 2019, it was reported that the device did not start up when plugged to mains.The blade and spacer were replaced the unit started up with no problem.Similar situation occurred number of times.Customer requested repair/checkup of the unit.The customer returned an electric dermatome device, serial number ((b)(4)), for evaluation.The device history record (dhr) review noted no related non-conformances, requests for deviation, change notices or any other issues with manufacturing or with the device.The dhr review also found that all verifications, inspections and tests were successfully completed.Product review of the electric dermatome by zimmer biomet on 05 november 2019 revealed that the reported event was confirmed as the motor speed was not steady, the unit was out of calibration, the needle bearing was worn, and the control bar was too thing at the edges.It was also noted that the device had non original screws installed at the neck.Repair of the electric dermatome was performed by zimmer biomet which included replacement of the needle bearing, shaft bearings, sleeve bearings, control bar, motor, and screws.Electric dermatome, serial number ((b)(4)), was then tested and functioned properly.It was repaired, inspected and tested.Although the reported event was confirmed during inspection of the device, it cannot be determined from the information provided what caused the components to fail.Therefore, the root cause could not be determined.The investigation was based on the information that was provided initially and any information that was obtained throughout the follow-up process.Review of the information provided during the investigation determined that there are no further actions needed at this time (ie/capa/scar/hhe/d).This complaint will be tracked and trended for any adverse trends that may warrant further action.
 
Event Description
This device was sent in for annual/preventative maintenance and there was no event.During evaluation of the device, it was revealed that the motor speed was not steady and the device did not have the original screws installed at the neck.No adverse events were reported as a result of this malfunction.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HANDPIECE, ELECTRIC DERMATOME
Type of Device
DERMATOME
Manufacturer (Section D)
ZIMMER SURGICAL, INC.
200 west ohio avenue
dover OH 44622
Manufacturer (Section G)
ZIMMER SURGICAL, INC.
200 west ohio avenue
dover OH 44622
Manufacturer Contact
christina arnt
56 e. bell drive
warsaw, IN 46582
5745273773
MDR Report Key9607915
MDR Text Key193634525
Report Number0001526350-2020-00079
Device Sequence Number1
Product Code GFD
Combination Product (y/n)N
Reporter Country CodePL
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 01/20/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/20/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number00882100100
Device Lot Number63733145
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/24/2019
Was the Report Sent to FDA? No
Date Manufacturer Received12/26/2019
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/27/2017
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Removal/Correction NumberN/A
Patient Sequence Number1
-
-