This follow-up report is being submitted to relay additional information.The following sections were updated: b4; b5; g4; g7; h1; h2; h3; h6.Visual examination of the provided pictures identified signs of repeated use (nicks and gouges) and is chipped/fractured.The reported event was confirmed.Device history record (dhr) was reviewed and no discrepancies were found.Per package insert, inspect all instruments/provisional carefully prior to each use.If damage or wear is noted that may compromise the function of the instrument, do not use the device and contact your zimmer representative for a replacement.The root cause of the reported issue was due to repeated use of the instrument for more than 15+ years field life; which causes wear and tear to the instrument and led to the fracture.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|